[Durgical](https://durgical.com/en/) › [DurgiPedia](https://durgical.com/en/durgipedia/) › [Anaesthesia and ICU](https://durgical.com/en/anaesthesia/) › [Critical care (ICU)](https://durgical.com/en/anaesthesia/areas/critical-care/) › Mechanical circulatory support: balloon pump, ECMO (VA and VV) and ventricular assist devices

# Mechanical circulatory support: balloon pump, ECMO (VA and VV) and ventricular assist devices

- Concept

When the heart or lungs fail despite drugs, machines can take over part of their work. Each machine helps in a different way and has its own costs.

Updated Nov 15, 2026

## Key points

- IABP: inflates in diastole (more coronary flow), deflates before systole (less afterload); modest help for a beating heart.
- VV ECMO replaces the lungs; VA ECMO supports the heart and the lungs.
- VA ECMO raises the afterload of the [left ventricle](https://durgical.com/en/anatomy/left-ventricle/): watch for distension and pulmonary oedema.
- Femoral VA ECMO with recovering heart and poor lungs: differential hypoxia; measure saturation from the right arm.
- Every device needs anticoagulation: bleeding, clotting, limb ischaemia, haemolysis and stroke.
- Support is a bridge: decide the destination (recovery, transplant, durable device or withdrawal) before starting.

On this page

- The basics Key points
- What it is
- How to think about it
- Key points
- Components
- How it works
- The devices compared: what each supports, when it is used and what it costs
- In practice
- Next step
- Common traps
- Elsewhere in DurgiPedia

- Study Test yourself

- Connections Connections

## What it is

Mechanical circulatory support means devices that assist or replace the pumping of the heart, the gas exchange of the lungs, or both, while the organ recovers or a longer-term decision is made. The intra-aortic balloon pump (IABP) is a balloon in the descending aorta that inflates in diastole and deflates just before systole; it gives modest help to a heart that is still beating. Extracorporeal membrane oxygenation (ECMO) drains blood from a large vein, pumps it through a membrane oxygenator and returns it: to a vein in venovenous (VV) ECMO, which replaces the lungs only, or to an artery in venoarterial (VA) ECMO, which supports both heart and lungs. Ventricular assist devices (VADs) take blood from a ventricle (usually the left) and pump it forward: percutaneous devices for days to weeks, and surgically implanted durable left ventricular assist devices (LVADs) for months to years, as a bridge to transplant or as permanent (destination) therapy. All need anticoagulation and specialist teams, and all carry bleeding, clotting, limb ischaemia, infection and stroke risks.

## How to think about it

Ask two things of every device: what does it replace, and where does it put the blood? The balloon pump replaces nothing; it only makes the heart's own beat easier, so it needs a working heart. VV ECMO replaces the lungs and returns blood to the venous side, so the heart must still pump. VA ECMO replaces heart and lungs but pushes blood back up the aorta against the left ventricle, so its price is a higher load on a failing ventricle. A VAD takes blood out of the ventricle and pushes it forward, so it unloads the ventricle.

## Key points

- IABP: a balloon in the descending aorta, timed to the ECG or arterial pressure; it needs a rhythm or pulse pressure to trigger.
- VV ECMO: blood drained from a large vein, oxygenated and returned to the venous side; used in severe respiratory failure such as [ARDS](https://durgical.com/en/conditions/acute-respiratory-distress-syndrome/) when the heart is adequate.
- VA ECMO: blood drained from a vein and returned into an artery; used in refractory [cardiogenic shock](https://durgical.com/en/conditions/cardiogenic-shock/), failure to come off cardiopulmonary bypass and, in selected centres, [cardiac arrest](https://durgical.com/en/conditions/cardiac-arrest/) (extracorporeal CPR).
- Percutaneous VADs (microaxial pumps placed across the aortic valve, or devices that take blood from the [left atrium](https://durgical.com/en/anatomy/left-atrium/)) give short-term support in cardiogenic shock and high-risk coronary intervention.
- Durable LVADs are implanted surgically for [advanced heart failure](https://durgical.com/en/conditions/advanced-heart-failure-transplant-and-lvad/) as a bridge to transplant or as destination therapy.
- Selection decides outcome: a reversible cause or a clear destination, the time without circulation, age and other organ failure.

## Components

- IABP: a balloon catheter inserted through the [femoral artery](https://durgical.com/en/anatomy/femoral-artery/) with its tip just below the left [subclavian artery](https://durgical.com/en/anatomy/right-subclavian-artery/), and a console that times inflation.
- ECMO circuit: a drainage cannula, a centrifugal pump, a membrane oxygenator with a gas supply, a heat exchanger and a return cannula.
- Peripheral VA ECMO: drainage from the [femoral vein](https://durgical.com/en/anatomy/femoral-vein/), return into the femoral artery, and often a small extra cannula that feeds the leg below the arterial cannula.
- Durable LVAD: an inflow cannula in the apex of the left ventricle, the pump, an outflow graft to the [ascending aorta](https://durgical.com/en/anatomy/ascending-aorta/) and a driveline through the skin to a controller and batteries.

## How it works

- Counterpulsation: inflation in diastole pushes blood back into the coronary arteries and forward to the body; sudden deflation just before systole leaves a space the ventricle ejects into more easily.
- Peripheral VA ECMO pumps blood up the aorta against the left ventricle; a very weak ventricle may fail to open the aortic valve, so it distends and the lungs flood unless it is vented.
- Differential hypoxia: with femoral VA ECMO, a recovering heart pumping blood through poor lungs sends poorly oxygenated blood to the coronaries and brain, while the legs receive well-oxygenated ECMO blood.
- Continuous-flow LVADs produce little or no pulse, so a cuff and a pulse oximeter may not read; a Doppler gives the mean pressure.

## The devices compared: what each supports, when it is used and what it costs

| Device | What it supports | When it is used | Main trade-off |
| --- | --- | --- | --- |
| [Intra-aortic balloon pump](https://durgical.com/en/conditions/cardiogenic-shock/) | Helps the heart's own beat: lower afterload, more coronary flow | Mechanical complications of infarction, bridge to surgery; routine use in infarct shock did not improve survival | Modest support; needs a rhythm; leg ischaemia; not with [aortic regurgitation](https://durgical.com/en/conditions/aortic-regurgitation/) or dissection |
| [VV ECMO](https://durgical.com/en/conditions/acute-respiratory-distress-syndrome/) | The lungs only (gas exchange) | Severe ARDS despite lung-protective ventilation and prone positioning | Needs a working heart; bleeding and clotting; specialist centres only |
| [VA ECMO](https://durgical.com/en/conditions/cardiogenic-shock/) | Heart and lungs | Refractory cardiogenic shock, failure to wean from bypass, cardiac arrest in selected patients | Raises left ventricular afterload; leg ischaemia; stroke; differential hypoxia |
| Percutaneous VAD | The left (or right) ventricle, short term | Cardiogenic shock, high-risk coronary intervention | Haemolysis, bleeding, vascular injury |
| [Durable LVAD](https://durgical.com/en/conditions/heart-failure/) | The left ventricle, months to years | Advanced [heart failure](https://durgical.com/en/conditions/heart-failure/): bridge to transplant or destination therapy | Driveline infection, pump thrombosis, stroke, gut bleeding, right ventricular failure |

## In practice

- Daily checks: cannula and driveline sites, perfusion of the leg below an arterial cannula, signs of haemolysis, anticoagulation, neurological state and the circuit.
- Gut bleeding is common with LVADs because the low pulse causes fragile gut vessels and the pump breaks down von Willebrand factor.
- An LVAD patient who collapses: check the patient and the device together (alarms, controller, batteries, driveline) and call the implanting team; follow their guidance on chest compressions.
- ECMO does not treat the disease: it buys time while the heart or lungs recover or a decision is made, so goals of care are discussed early with the family.
- Right ventricular failure can appear after an LVAD starts, because the [right ventricle](https://durgical.com/en/anatomy/right-ventricle/) must now fill a left ventricle that is emptied faster.

## Next step

- [Refractory shock](https://durgical.com/en/conditions/refractory-shock/) or hypoxaemia despite optimal treatment: call the critical care and cardiac or ECMO team early, before multiorgan failure makes support futile.
- In Colombia few centres offer ECMO: discuss transfer early, while the patient can still travel safely.
- On VA ECMO, check right-arm saturation and arterial gases for differential hypoxia and an echocardiogram for left ventricular distension.
- For any LVAD patient, contact the implanting centre for advice and never disconnect the driveline.
- Agree the bridge and the exit strategy (recovery, transplant, durable device or withdrawal) and document the goals of care.

## Common traps

- **Mistaken belief**: VV ECMO supports a failing heart. Correction: it returns blood to the veins and replaces only gas exchange; the patient's own heart must still pump.
- **Mistaken belief**: the balloon pump is standard for shock after a heart attack. Correction: routine use did not improve survival; it is kept for selected cases such as mechanical complications or a bridge to surgery.
- **Mistaken belief**: a good saturation from the leg means the brain is well oxygenated on femoral VA ECMO. Correction: the heart may be sending poorly oxygenated blood to the head; measure from the right arm.
- **Mistaken belief**: an LVAD patient without a palpable pulse is in cardiac arrest. Correction: continuous-flow pumps often give no pulse; check consciousness, perfusion, the device and a Doppler pressure.

## Test yourself

Think of the answer, then tap the question to see it.

## When does an intra-aortic balloon pump inflate and deflate, and why?

It inflates in diastole, raising coronary perfusion pressure, and deflates just before systole, lowering the afterload the left ventricle ejects against.

## What is the key difference between VV and VA ECMO?

VV ECMO returns oxygenated blood to a vein and replaces only the lungs; VA ECMO returns it to an artery and supports both the circulation and gas exchange.

## Why can peripheral VA ECMO make the left ventricle distend?

It pumps blood up the aorta against the ventricle, raising its afterload; a very weak ventricle cannot open the aortic valve, so blood accumulates and the lungs flood.

## What is differential hypoxia on femoral VA ECMO, and where is saturation measured?

A recovering heart ejects poorly oxygenated blood from bad lungs to the coronaries and brain while the legs get ECMO blood; saturation is measured from the right arm.

## Why do patients with a continuous-flow LVAD often bleed from the gut?

The low pulse pressure favours fragile abnormal gut vessels, the pump breaks down von Willebrand factor, and they are anticoagulated.

[Add to my cards (5)](https://durgical.app/estudiar/?add=/en/anaesthesia/mechanical-circulatory-support-balloon-pump-ecmo-and-ventricular-assist-devices/)

## Elsewhere in DurgiPedia

- [Preload, afterload and contractility](https://durgical.com/en/physiology/preload-afterload-and-contractility/)
- [Cardiac output and what sets it](https://durgical.com/en/physiology/cardiac-output-and-what-sets-it/)
- [Shock: the clinical signs and the types](https://durgical.com/en/clinical-exam/shock-the-clinical-signs-and-the-types/)

## Connections

## Related 7

- [Shock: a bedside approach by type](https://durgical.com/en/anaesthesia/shock-a-bedside-approach-by-type/)
- [Shock haemodynamic profiles: output, resistance, filling pressures and venous saturation](https://durgical.com/en/anaesthesia/shock-haemodynamic-profiles-output-resistance-filling-pressures-and-venous-satur/)
- [Vasopressors and inotropes by receptor and mechanism](https://durgical.com/en/anaesthesia/vasopressors-and-inotropes-by-receptor-and-mechanism/)
- [Ventilating ARDS: low volumes, prone position and more](https://durgical.com/en/anaesthesia/ventilating-ards-low-volumes-prone-position-and-more/)
- [Anaesthesia for cardiac surgery, in principle](https://durgical.com/en/anaesthesia/anaesthesia-for-cardiac-surgery-in-principle/)
- [Care after cardiac arrest](https://durgical.com/en/anaesthesia/care-after-cardiac-arrest/)
- [Haemodynamic monitoring: arterial line, CVP, cardiac output and lactate](https://durgical.com/en/anaesthesia/haemodynamic-monitoring-arterial-line-cvp-cardiac-output-and-lactate/)

## Related conditions 5

- [Cardiogenic shock](https://durgical.com/en/conditions/cardiogenic-shock/)
- [Acute respiratory distress syndrome (ARDS)](https://durgical.com/en/conditions/acute-respiratory-distress-syndrome/)
- [Heart failure](https://durgical.com/en/conditions/heart-failure/)
- [Myocarditis](https://durgical.com/en/conditions/myocarditis/)
- [Pulmonary embolism](https://durgical.com/en/conditions/pulmonary-embolism/)

[← All of anaesthesia](https://durgical.com/en/anaesthesia/)

Educational content; it does not replace supervised training, your institution's protocols or the anaesthetist's or intensivist's decision. It contains no doses and no ventilator or infusion settings.

Sources: [Intra-Aortic Balloon Pump](https://www.ncbi.nlm.nih.gov/books/NBK542233/) (StatPearls (NCBI Bookshelf), 2023); [Extracorporeal Membrane Oxygenation in Adults](https://www.ncbi.nlm.nih.gov/books/NBK576426/) (StatPearls (NCBI Bookshelf), 2025); [Percutaneously Inserted Ventricular Assist Device](https://www.ncbi.nlm.nih.gov/books/NBK554379/) (StatPearls (NCBI Bookshelf), 2026); [EMS Left Ventricular Assist Device Management](https://www.ncbi.nlm.nih.gov/books/NBK525987/) (StatPearls (NCBI Bookshelf), 2023); [Cardiogenic Shock](https://www.ncbi.nlm.nih.gov/books/NBK482255/) (StatPearls (NCBI Bookshelf), 2023).

## **Spotted a mistake?** We are always improving for the community. Tell us what is wrong and our team will review it. Report a mistake

Representative model, for educational purposes only.

---
Source: https://durgical.com/en/anaesthesia/mechanical-circulatory-support-balloon-pump-ecmo-and-ventricular-assist-devices/
License: CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/). Cite as: DurgiPedia (durgical.com), https://durgical.com/en/anaesthesia/mechanical-circulatory-support-balloon-pump-ecmo-and-ventricular-assist-devices/