# Capnography (end-tidal CO2): tube confirmation, the waveform and what a change means

- Assessment

Capnography shows exhaled CO2 breath by breath: it confirms the tube, warns of disconnection and reflects ventilation and blood flow to the lungs.

Updated Nov 13, 2026

## Key points

- A persistent CO2 waveform after intubation means the tube is in the trachea; CO2 can appear for a few breaths when the tube is in the oesophagus.
- In normal lungs EtCO2 is about 3–5 mmHg lower than PaCO2; the gap widens with low blood pressure, pulmonary embolism, high PEEP, severe ARDS or emphysema.
- If the waveform suddenly disappears, check the patient, tube and circuit before the monitor.
- During CPR, an abrupt rise in EtCO2 is the first sign of return of spontaneous circulation.
- A long, sloping rise in the expiratory part of the curve suggests bronchospasm or airway obstruction.

## What it is

Capnometry measures the carbon dioxide at the end of each exhaled breath (EtCO2) as a number; capnography adds the waveform, CO2 over time. Mainstream sensors sit in line at the endotracheal tube and are used in intubated patients; sidestream systems draw a gas sample and also work in non-intubated patients through a nasal or oral-nasal cannula. Colorimetric detectors only change colour and give a range, not a value. Capnography confirms tube placement, monitors circuit continuity and the adequacy of ventilation, and during CPR reflects blood flow produced by chest compressions.

## Reference values

- Normal waveform: a square shape with a flat top (plateau).
- Phase 0: inspiration; the CO2 falls to zero.
- Phase I: start of expiration, dead-space gas with no CO2.
- Phase II: a rapid rise as mixed gas is exhaled.
- Phase III: the plateau of alveolar gas; its end value is the EtCO2.
- There is no single target: the team compares EtCO2 with the patient's own PaCO2.

## What you need

- Mainstream sensor in line at the endotracheal tube, or a sidestream sampling line.
- For non-intubated patients, a nasal or oral-nasal sampling cannula.
- A monitor that shows both the number and the waveform.
- A colorimetric detector where capnography is not available.

## How to measure it (and why)

- Choose mainstream for an intubated patient, or sidestream (also by cannula for a non-intubated one), as the protocol says.
- Fit a mainstream sensor in line at the tube hub, as the protocol says — it is designed for intubated patients.
- After intubation, look for a waveform that persists over about six breaths — a few early breaths can come from the oesophagus.
- Check the shape every time: a square wave with a plateau — without a plateau the number is unreliable.
- Compare EtCO2 with the PaCO2 on the blood gas and note the gap — EtCO2 underestimates PaCO2 when lung blood flow falls.
- Watch the trend with ventilator changes, sedation and transport — rising EtCO2 can mean hypoventilation; falling can mean low output.
- If the waveform disappears, check the patient, the tube and the circuit first — disconnection or displacement must be ruled out.
- During CPR, use EtCO2 to judge compressions and watch for an abrupt rise — it reflects cardiac output and return of circulation.
- Record the value, the waveform and the context with each check — the trend matters more than one number.

## How to read it

- Low EtCO2 with low blood pressure: reduced blood flow to the lungs, as in shock.
- Low EtCO2 with a wide gap to PaCO2: more dead space, as in pulmonary embolism.
- A prolonged, sloping expiratory phase: bronchospasm, as in COPD or asthma.
- No CO2 after intubation: the tube is in the oesophagus until proven otherwise.
- Colorimetric detectors can give a false positive after bag-mask ventilation or recent carbonated drinks or antacids; the reading is usually reliable after about six breaths. Sodium bicarbonate raises EtCO2 for 5–10 minutes.

## Next step: when to call for help

- No waveform after intubation, or a waveform that disappears: tell the provider or call for help at once.
- A sudden fall in EtCO2 with low blood pressure: report it now.
- EtCO2 rising with drowsiness or sedation: report it; the patient may be hypoventilating.
- A new change in the waveform shape: report it and check the airway.

## Test yourself

Think of the answer, then tap the question to see it.

## Why is a CO2 waveform watched over several breaths after intubation?

Because a little CO2 can appear for a few breaths even when the tube is in the oesophagus; a waveform that persists confirms the trachea.

## In normal lungs, how does EtCO2 compare with PaCO2?

EtCO2 is about 3–5 mmHg lower. The gap widens when lung blood flow falls or dead space increases.

## What does an abrupt rise in EtCO2 during CPR suggest?

Return of spontaneous circulation; it is often the first sign.

## What is the first thing to do if the capnography waveform suddenly disappears?

Check the patient, the tube and the circuit for disconnection, displacement or obstruction before blaming the monitor.

[Add to my cards (4)](https://durgical.app/estudiar/?add=/en/nursing/capnography-tube-confirmation-the-waveform-and-what-a-change-means/)

## Elsewhere in DurgiPedia

- [Dead space and alveolar ventilation](https://durgical.com/en/physiology/dead-space-and-alveolar-ventilation/)

## Connections

## Related 3

- [Assisting endotracheal intubation: preparing, rapid sequence from the nursing side and confirming tube position](https://durgical.com/en/nursing/assisting-endotracheal-intubation-preparing-rapid-sequence-from-the-nursing-side/)
- [Pulse oximetry at the bedside: placement and pitfalls](https://durgical.com/en/nursing/pulse-oximetry-at-the-bedside-placement-and-pitfalls/)
- [Basic life support from the nursing side](https://durgical.com/en/nursing/basic-life-support-from-the-nursing-side/)

[← All of nursing](https://durgical.com/en/nursing/)

Educational content for students; it does not replace your institution's protocols, supervised clinical training or a prescriber's order. Dose-calculation examples use made-up numbers: they are educational, not a clinical order.

Sources: [Clinical care for severe acute respiratory infection: toolkit, update 2022 (COVID-19 adaptation), tools 4.4 and 9.4](https://www.who.int/publications/i/item/clinical-care-of-severe-acute-respiratory-infections-tool-kit) (World Health Organization, 2022); [Applications of end-tidal carbon dioxide (ETCO2) monitoring in emergency department; a narrative review](https://pmc.ncbi.nlm.nih.gov/articles/PMC5827051/) (Emergency (Tehran) (open access, PMC), 2018); [Use of capnography during resuscitation of patients in intensive care units](https://pmc.ncbi.nlm.nih.gov/articles/PMC12126665/) (SAGE Open Nursing (open access, PMC), 2025).

## **Spotted a mistake?** We are always improving for the community. Tell us what is wrong and our team will review it. Report a mistake

Representative model, for educational purposes only.

---
Source: https://durgical.com/en/nursing/capnography-tube-confirmation-the-waveform-and-what-a-change-means/
License: CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/). Cite as: DurgiPedia (durgical.com), https://durgical.com/en/nursing/capnography-tube-confirmation-the-waveform-and-what-a-change-means/