# Cardiac monitoring at the bedside: leads, alarms and rhythm changes to report

- Safe practice

Continuous monitoring shows the heart rhythm only if electrodes are placed well and every alarm is checked at the patient before it is silenced.

Updated Nov 12, 2026

## Key points

- 5-lead: RA and LA below the right and left collarbones, RL and LL on the right and left lower chest below the ribs, V1 at the 4th intercostal space, right sternal border.
- For a 3-lead system, follow the electrode labels and the protocol.
- Connect each wire by the letter and colour printed on it, as the protocol says.
- Never silence or switch off an alarm without looking at the patient and the rhythm.
- Report at once: ventricular tachycardia or fibrillation, asystole, complete heart block, new irregular rhythm, ST changes.

## What it is

Bedside or telemetry monitoring records one or more ECG leads continuously so that dangerous rhythm changes are seen in time. Its safety depends on three things: correct electrode placement, alarms that are switched on and set for the patient, and a nurse who answers each alarm by looking at the patient first. Many alarms are false (artefact, a loose electrode) or do not need action, and that leads to alarm fatigue; the answer is better skin preparation, regular electrode changes and alarm limits adjusted to the patient, never silencing without checking.

## What it requires

- A medical order and a unit protocol for who watches the monitor and how alarms are answered.
- Electrodes on clean, dry, hair-free skin, changed as the protocol says (often daily).
- Alarms switched on and audible, with limits set for the patient and their order.
- A rhythm strip printed and interpreted at the start of the shift and with every change.

## Components

- Electrodes, lead wires and the bedside or telemetry monitor.
- Alarm types: rhythm alarms, rate limits and technical alarms such as a lead off.
- False alarms (invalid signal) differ from non-actionable alarms (accurate but not clinically important); each needs a different fix.
- Rhythm strip: usually a 6-second strip to estimate rate and rhythm.

## How it is applied (and why)

- Check the order, identify the patient, explain the purpose and ensure privacy — consent and fewer interrupted readings.
- Prepare the skin: clean, dry and hair-free, with electrodes stuck firmly — good contact reduces false alarms.
- Place the electrodes for the lead system in use and connect them by their letters — misplacement distorts the rhythm.
- Choose a lead with clear P waves and QRS complexes, check the rhythm and print a strip — a baseline to compare.
- Set alarm limits for the patient and the order, and check they are audible — default limits may not fit.
- At every alarm, look at the patient first: responsive, breathing, pulse; then the rhythm — a flat line may be a real arrest.
- Fix technical causes such as a loose electrode or lead off, and change electrodes as scheduled — fewer false alarms.
- Record and report strips at handoff and with any change — the team can see trends.

## In nursing practice

- Any rhythm with an unresponsive patient and no pulse: call the emergency team and start resuscitation.
- Ventricular tachycardia: fast, wide, regular complexes; check the pulse and call at once.
- Ventricular fibrillation: a chaotic line with no clear complexes; the patient is in cardiac arrest.
- Asystole: a flat line; check the patient and leads in seconds, and resuscitate if unresponsive.
- Complete heart block: P waves and QRS complexes march independently; it can lead to arrest, so report at once.
- New atrial fibrillation: irregularly irregular rhythm without clear P waves; assess blood pressure and signs of stroke and report.
- New ST elevation or depression, especially with chest pain: urgent.
- Heart rate very slow or very fast with symptoms (dizziness, low blood pressure, chest pain, confusion).

## Common mistakes

- Silencing an alarm from the station without seeing the patient.
- Leaving default alarm limits that do not fit the patient.
- Connecting wires without checking their labels.
- Leaving dry, old electrodes that cause repeated false alarms.
- Assuming a flat line is a loose lead without checking the patient.

## Test yourself

Think of the answer, then tap the question to see it.

## What is the first thing to do when a cardiac monitor alarms?

Look at the patient: responsiveness, breathing and pulse. Only then interpret the rhythm and look for technical causes.

## Where is V1 placed in a 5-lead system?

In the 4th intercostal space at the right sternal border.

## Which two measures reduce false monitor alarms?

Good skin preparation (clean, dry, hair-free) and changing the electrodes as the protocol says, usually daily.

## What does ventricular fibrillation look like on the monitor?

A chaotic, wavy line with no clear complexes. The patient has no effective heartbeat and is in cardiac arrest.

[Add to my cards (4)](https://durgical.app/estudiar/?add=/en/nursing/cardiac-monitoring-at-the-bedside-leads-alarms-and-rhythm-changes-to-report/)

## Connections

## Related 4

- [Recording a 12-lead ECG](https://durgical.com/en/nursing/recording-a-12-lead-ecg/)
- [Recognising deterioration and calling the rapid response team](https://durgical.com/en/nursing/recognising-deterioration-and-calling-the-rapid-response-team/)
- [Basic life support from the nursing side](https://durgical.com/en/nursing/basic-life-support-from-the-nursing-side/)
- [The crash cart: checks and organisation](https://durgical.com/en/nursing/the-crash-cart-checks-and-organisation/)

[← All of nursing](https://durgical.com/en/nursing/)

Educational content for students; it does not replace your institution's protocols, supervised clinical training or a prescriber's order. Dose-calculation examples use made-up numbers: they are educational, not a clinical order.

Sources: [Nursing Advanced Skills, 7.6: Checklist - Initiate Telemetry](https://med.libretexts.org/Bookshelves/Nursing/Nursing_Advanced_Skills_(OpenRN)/07%3A_Interpret_Basic_ECG/7.06%3A_Checklist-_Initiate_Telemetry) (Open RN (Chippewa Valley Technical College), LibreTexts, 2023); [Nursing Advanced Skills, 7.4: ECG Patterns and Dysrhythmias](https://med.libretexts.org/Bookshelves/Nursing/Nursing_Advanced_Skills_(OpenRN)/07%3A_Interpret_Basic_ECG/7.04%3A_ECG_Patterns_and_Dysrhythmias) (Open RN (Chippewa Valley Technical College), LibreTexts, 2023); [Testing physiologic monitor alarm customization software to reduce alarm rates and improve nurses' experience of alarms in a medical intensive care unit (Ruppel H et al.)](https://pmc.ncbi.nlm.nih.gov/articles/PMC6193710/) (PLoS One (open access, PMC), 2018).

## **Spotted a mistake?** We are always improving for the community. Tell us what is wrong and our team will review it. Report a mistake

Representative model, for educational purposes only.

---
Source: https://durgical.com/en/nursing/cardiac-monitoring-at-the-bedside-leads-alarms-and-rhythm-changes-to-report/
License: CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/). Cite as: DurgiPedia (durgical.com), https://durgical.com/en/nursing/cardiac-monitoring-at-the-bedside-leads-alarms-and-rhythm-changes-to-report/