# Temporary pacing (transcutaneous and transvenous): set-up, capture, sensing and safety

- Procedure

Temporary pacing keeps the heart rate up until the cause is treated; the nurse checks capture on the monitor and at the pulse, comfort, skin and settings.

Updated Nov 13, 2026

## Key points

- Electrical capture: a pacing spike followed by a wide QRS and a T wave; do not take the spike's afterpotential for capture.
- Mechanical capture: a pulse that matches the paced rate; check it at the right femoral or carotid artery so muscle twitching does not fool you.
- Transcutaneous pacing hurts: sedation and analgesia as ordered, and reassess comfort often.
- Demand (synchronous) mode is the usual mode; fixed-rate (asynchronous) pacing risks an R-on-T arrhythmia.
- Undersensing can trigger ventricular tachycardia or fibrillation; keep a defibrillator ready.

## What it is

Temporary pacing gives electrical impulses to the heart when its own rate is too slow or unreliable. Transcutaneous pacing uses pads on the chest and back; it is quick to start but painful and only a bridge. Transvenous pacing uses a wire passed through a central vein into the right ventricle; epicardial pacing uses wires placed on the heart during surgery. The nurse prepares the equipment, places the pads correctly, confirms electrical capture on the ECG and mechanical capture at a pulse, treats pain and anxiety as ordered, checks the skin, connections and settings, and knows what to do when the pacemaker fails to capture or to sense.

## Indications

- Symptomatic or unstable slow heart rhythms that do not respond to drugs, such as high-grade heart block, as the medical team decides.
- A bridge to transvenous or permanent pacing.
- After heart surgery, through epicardial wires, to treat slow rhythms or keep AV timing.

## Contraindications and precautions

- Transcutaneous pacing is relatively contraindicated in hypothermia, and is not useful in prolonged asystolic arrest.
- It is a temporary bridge only: thresholds rise with time, and in some patients tolerable capture cannot be achieved.
- Higher pacing thresholds are expected after chest surgery, with emphysema, pericardial effusion or positive pressure ventilation.
- Clean, dry, hair-trimmed skin under the pads; no alcohol or flammable liquids.

## Equipment

- Monitor-defibrillator with pacing function and pacing pads (transcutaneous).
- Pacing box with fresh battery and connecting cables (transvenous or epicardial).
- Central venous access set and sterile barrier for transvenous insertion, as the protocol lists.
- ECG monitoring, emergency drugs and the crash cart nearby.

## Step by step (and why)

- Explain the procedure and its discomfort, and give sedation and analgesia as ordered — pain and anxiety make pacing hard to tolerate.
- Trim hair, wash with soap and water and dry the skin — poor contact causes failure to capture, pain and burns.
- Place the front pad over the apex or at V3 and the back pad left of the spine behind it — no bone between the pads and the heart.
- Set the rate and mode as ordered, usually demand mode — fixed-rate pacing can fire on the T wave.
- Raise the output until each spike is followed by a wide QRS and a T wave, then set it above that threshold as ordered — this keeps capture reliable.
- Check a pulse that matches the paced rate at the right femoral or carotid artery — this proves the heart is pumping, not just firing.
- Transvenous or epicardial: check connections, battery, settings and which wire is which — loose connections are a common cause of failure.
- Check the skin under the pads often and reposition as needed — prolonged pacing can burn the skin.
- Record mode, rate, output, sensitivity, threshold and the patient's response — the next nurse needs the same baseline.

## Complications

- Failure to capture: wire or pad movement, poor contact, loose connections, a low battery, effusion, ischaemia or acidosis.
- Oversensing (for example from shivering) can stop the pacemaker from firing; undersensing can fire on the T wave.
- Hiccups or coughing from diaphragm stimulation, and pain from muscle contraction.
- Skin burns under pads, especially with long pacing.
- Transvenous wire complications: misplacement, pneumothorax, pericardial effusion or tamponade.

## Next step: when to call for help

- Call at once if capture is lost, the pulse does not match the paced rate, or the patient becomes hypotensive or loses consciousness.
- Call for a new arrhythmia, especially ventricular tachycardia or fibrillation, and be ready to defibrillate.
- Report rising thresholds, pain that does not settle with the ordered treatment, or skin damage under the pads.
- After transvenous insertion, report breathlessness, chest pain or falling pressure: pneumothorax or tamponade must be ruled out.

## Test yourself

Think of the answer, then tap the question to see it.

## How is electrical capture confirmed during transcutaneous pacing?

Each pacing spike is followed by a wide QRS with a distinct ST segment and T wave; the afterpotential after the spike is not capture.

## Where is the pulse checked to confirm mechanical capture, and why?

At the right femoral or carotid artery, so that muscle twitching from the pacing current is not mistaken for a pulse.

## Why is demand mode preferred to fixed-rate pacing?

Fixed-rate pacing ignores the heart's own beats and can fire on the T wave, which may trigger ventricular tachycardia or fibrillation.

## Which skin care reduces pain and failure of transcutaneous pacing?

Trimming (not shaving) hair, washing with soap and water, drying the skin well and avoiding alcohol under the pads.

[Add to my cards (4)](https://durgical.app/estudiar/?add=/en/nursing/temporary-pacing-set-up-capture-sensing-and-safety/)

## Connections

## Related 3

- [Cardiac monitoring at the bedside: leads, alarms and rhythm changes to report](https://durgical.com/en/nursing/cardiac-monitoring-at-the-bedside-leads-alarms-and-rhythm-changes-to-report/)
- [The crash cart: checks and organisation](https://durgical.com/en/nursing/the-crash-cart-checks-and-organisation/)
- [After heart surgery in the ICU: bleeding, tamponade, rewarming, pacing wires and chest drains](https://durgical.com/en/nursing/after-heart-surgery-in-the-icu-bleeding-tamponade-rewarming-pacing-wires-and-che/)

[← All of nursing](https://durgical.com/en/nursing/)

Educational content for students; it does not replace your institution's protocols, supervised clinical training or a prescriber's order. Dose-calculation examples use made-up numbers: they are educational, not a clinical order.

Sources: [Using transcutaneous cardiac pacing to best advantage: How to ensure successful capture and avoid complications (Doukky R et al., J Crit Illn 2003;18(5):219-225)](https://pmc.ncbi.nlm.nih.gov/articles/PMC6376978/) (Journal of Critical Illness, open access in PubMed Central, 2003); [Temporary epicardial pacing after cardiac surgery (Gillham MJ, Barr TM, BJA Education 2023;23(9):337-349)](https://pmc.ncbi.nlm.nih.gov/articles/PMC10433321/) (BJA Education, open access in PubMed Central, 2023); [Troubleshooting Transvenous Pacemakers with Point of Care Ultrasound (POCUS) (Gohde E et al., POCUS J 2025;10(1):53-60)](https://pmc.ncbi.nlm.nih.gov/articles/PMC12057482/) (POCUS Journal, open access in PubMed Central, 2025).

## **Spotted a mistake?** We are always improving for the community. Tell us what is wrong and our team will review it. Report a mistake

Representative model, for educational purposes only.

---
Source: https://durgical.com/en/nursing/temporary-pacing-set-up-capture-sensing-and-safety/
License: CC BY 4.0 (https://creativecommons.org/licenses/by/4.0/). Cite as: DurgiPedia (durgical.com), https://durgical.com/en/nursing/temporary-pacing-set-up-capture-sensing-and-safety/