Measurement variation

  • Concept

Measuring the same thing twice almost never gives the same result: the person changes, the observer changes and the device changes.

Updated

What it is

Every clinical measurement varies for several reasons: the person's biological variation from one moment to the next (within-subject), the same observer repeating a reading (intra-observer), different observers (inter-observer) and the instrument. Knowing these sources helps decide how many measurements to take, how to standardise them and when a change between two visits is real. This page covers the sources of variation and how to reduce them; the difference between validity and reliability has its own page.

Worked example

  1. A patient's systolic pressure: the standard deviation between readings in the same person is 6 mmHg.
  2. With a single reading, 95% of values fall within ± 1.96 × 6 ≈ ± 12 mmHg of the usual value.
  3. With the mean of 3 readings: 6 / √3 ≈ 3.5 mmHg; interval ± 1.96 × 3.5 ≈ ± 7 mmHg.
  4. With 4 readings: 6 / √4 = 3 mmHg; interval ± 5.9 mmHg.
  5. A reading of 146 mmHg with a usual value of 136 fits within the variation of a single measurement.
  6. Reading: averaging several readings cuts the noise nearly in half; deciding on one reading can misclassify.

Sources of variation and how to reduce them

SourceExampleHow to reduce it
Within-subject (biological)Blood pressure changing during the daySeveral measurements, same time and conditions
Intra-observerSame radiologist, different reading another dayWritten criteria, reading blind to the previous one
Inter-observerTwo doctors hear different soundsJoint training, operational definitions, measure kappa or ICC
InstrumentAn uncalibrated blood pressure monitorRegular calibration against a standard

Markers

  • Intra-observer
  • Inter-observer
  • Kappa
  • ICC

In practice

  • Before diagnosing on a single value (blood pressure, glucose, spirometry), confirm it with repeated measurements.
  • A change between two visits is real only if it exceeds the expected measurement variation.
  • In studies, random error in the exposure usually weakens the association toward the null.

Common pitfalls

  • Believing that averaging fixes an uncalibrated device: it only reduces random error.
  • Comparing studies that measured with different protocols.
  • Not reporting the reliability (kappa, ICC) of the main measurements.

Test yourself

Think of the answer, then tap the question to see it.

Within-subject SD of 6 mmHg. What is the standard error of the mean of 4 readings?

6 / √4 = 3 mmHg.

Which type of error does averaging reduce, and which does it not?

It reduces random error; it does not fix systematic error (bias), which needs calibration.

How is inter-observer agreement measured for a categorical variable?

With the kappa statistic, which corrects for agreement expected by chance.

Add to my cards (3)

Connections

← All of epidemiology

Educational content on epidemiological and statistical methods; the examples use made-up numbers to teach the calculation. Not a clinical or legal guide: current Colombian rules are in the official sources.