Checking that sterilisation worked: indicators
- Procedure
Each cycle is checked three ways: the machine's readings, chemical indicators that change colour, and regular spore tests proving microbes were killed.
What it is
Sterilisation can fail silently because of overloading, poor packaging, wrong cycles or machine faults. Monitoring combines physical indicators (time, temperature and pressure displayed or printed by the steriliser), chemical indicators (strips or inks inside and outside packs that change when conditions are met) and biological indicators (spore tests, the only direct proof that the cycle kills highly resistant organisms). Records allow instruments to be traced and recalled if a cycle fails.
Indications and contraindications
- Physical: read and record the cycle parameters every time.
- Chemical: an external indicator shows the pack was processed; an internal one shows steam reached the inside.
- Biological: spore tests at least weekly and after repairs or changes, with a control.
- Air-removal tests check pre-vacuum autoclaves daily.
- A positive spore test means the steriliser is taken out of use and retested at once; items are recalled and reprocessed as the protocol sets, implantable items always.
How it is done
- Check physical readings at the end of each cycle and record them.
- Inspect chemical indicators before using any pack.
- Run a biological indicator on schedule and send or incubate it with a control.
- Log results with the cycle number and load.
- Act on any failure: stop use, recall items, repair and retest.
| Indicator | What it shows | How often | What it cannot tell you |
|---|---|---|---|
| Physical (time, temperature, pressure) | The chamber reached the cycle settings | Every cycle | Whether steam reached inside each pack |
| External chemical (tape, pouch ink) | The pack went through a process | Every pack whose internal indicator cannot be seen | Whether the inside was sterilised |
| Internal chemical (strip inside the pack) | Conditions reached the instruments | Every pack | Whether organisms were killed |
| Biological (spore test) | Highly resistant spores were killed | At least weekly, and after repairs or changes | The result is delayed, so loads are traced by cycle |
Complications and pitfalls
- A colour change on a pouch does not prove sterility; it shows the pack saw the process.
- Staff need training to read indicators correctly.
- Keeping records protects patients and the practice in audits and investigations.
Next step / Initial management
- Keep a sterilisation log with physical, chemical and biological results.
- Failed spore test: take the steriliser out of use, recall and reprocess instruments, refer it for repair and report under the institution's protocol.
- Service and validate sterilisers on the manufacturer's schedule.
Common traps
- A changed colour on the pouch proves sterility: it shows only that the pack went through a process
- A good printout means every pack is sterile: overloading or poor packing can stop steam reaching some instruments
- A positive spore test can be ignored if the next load looks fine: take the steriliser out of use, review the procedure, retest at once and recall items as the protocol sets
- Spore tests are only for hospital sterilisers: dental sterilisers need them on schedule too
Test yourself
Think of the answer, then tap the question to see it.
Which indicator gives direct proof that a sterilisation cycle kills microorganisms?
A biological indicator (spore test).
What does a colour change on the outside of a pouch show?
Only that the pack went through a sterilisation process, not that it is sterile.
What must happen after a positive spore test?
The steriliser is taken out of use, its loading and cycle are reviewed and it is retested at once; items are recalled as the protocol sets, and it is repaired and retested if it fails again.
A 30-year-old nurse in a dental clinic gets a positive weekly spore test. What happens next?
Take the steriliser out of use, review how it was loaded and run, retest at once and recall items as the protocol sets (implantable items always); repair and retest if it fails again.
A patient's surgical pack is opened and the internal indicator has not changed. What do you do?
Do not use it: steam may not have reached the instruments; reprocess them, check the load and the cycle, and record the event.
Connections
Educational content; it does not replace assessment by a dentist or doctor and is not a clinical instruction. For facial or neck swelling with difficulty swallowing or breathing, a knocked-out tooth or bleeding that will not stop, go to an emergency department at once.
Sources: Decontamination and reprocessing of medical devices for health-care facilities (World Health Organization, 2016); Dental Infection Control (StatPearls Publishing (NCBI Bookshelf), 2023); Best Practices for Sterilization Monitoring in Dental Settings (Centers for Disease Control and Prevention, 2024).
Spotted a mistake?We are always improving for the community. Tell us what is wrong and our team will review it.Report a mistake