Operating-room fire and electrical safety
- Concept
Theatres mix oxygen, flammable drapes and sparks, and wire an unconscious patient to electrical equipment. Fires and electrical burns are preventable.
What it is
A surgical fire needs the three sides of the fire triangle at the same place and time: an oxidiser (oxygen or nitrous oxide), an ignition source (electrosurgery, laser, fibreoptic light cables, high-speed burrs) and a fuel (drapes, gauze, alcohol-based skin preparation, hair, the tracheal tube). Most fires happen on the head, neck and upper chest of a sedated patient breathing open oxygen under drapes; airway fires happen during laser or electrosurgery in the airway. Electrical safety concerns current flowing through the patient. In monopolar electrosurgery high-frequency current runs from the small active tip through the body to a large return pad and back to the generator; bipolar electrosurgery keeps it between the two tips of the forceps. Burns occur where current leaves through a small area: a poorly attached pad, an ECG electrode or metal touching the skin. Macroshock is current entering through the skin; microshock is a tiny current delivered straight to the heart through a pacing wire or a fluid-filled catheter, which can cause ventricular fibrillation. Implanted pacemakers, defibrillators and neurostimulators can be disturbed by electrosurgery.
Key points
- High-risk procedures are those with an ignition source near an oxidiser-rich area: face, head and neck surgery under sedation, and laser or electrosurgery in the airway.
- Under drapes, oxygen builds up near the face: use the lowest oxygen that keeps saturation safe, or air, and secure the airway if more oxygen is needed.
- Alcohol-based skin preparation must dry fully and must not pool in hair, skin folds or under drapes before an ignition source is used.
- Oxygen-enriched air makes materials ignite more easily and burn faster and hotter, and it can reach the field under drapes without being noticed.
- Monopolar electrosurgery passes current through the patient to a return pad; bipolar keeps it between the forceps tips and is safer near devices and in small areas.
- Microshock bypasses the protection of the skin, so a current far too small to feel can cause ventricular fibrillation.
Components
- The fire risk check at the time-out: is an ignition source going to be used near an oxidiser? If so, the team agrees how each side will be reduced and who will call stop.
- The return electrode: on clean, dry, hair-free skin over well-perfused muscle close to the site, away from bony points, metal implants and ECG electrodes.
- Equipment grounding, isolated power supplies and line isolation monitors in some theatres reduce the chance that a single fault shocks the patient or staff.
- Implanted devices: pacemakers, defibrillators and neurostimulators need a plan made with the device team, and bipolar electrosurgery or short bursts away from the device.
How it works
- Electrosurgery uses high-frequency current, which heats tissue without stimulating nerves and muscles; at the tiny active tip the current is concentrated and cuts, at the large pad it is spread and harmless.
- If the pad partly lifts, the same current crosses a smaller area and burns the skin; if the current finds another exit, such as an ECG electrode or a metal table, it burns there.
- Electrosurgery can be sensed by a pacemaker as heartbeats, so it stops pacing, or by a defibrillator as ventricular fibrillation, so it shocks.
Step by step
- Prevent: name oxidiser, fuel and ignition at the time-out and agree the oxygen plan before the first incision.
- Airway fire: stop the gas flow and remove the tracheal tube at once, because the burning tube is fuel inside the airway.
- Pour saline into the airway, then ventilate by mask with air, avoiding extra oxygen while anything still smoulders.
- Re-intubate, inspect the airway with a bronchoscope and remove debris, because airway burns can swell over hours.
- Fire on or around the patient: stop the gas, remove burning drapes and put it out with saline or water; use the extinguisher if it persists.
- Afterwards: examine the burns, keep the materials and report the event, because injury can be deeper than it first looks.
The fire triangle and the electrical path: who controls each, and how to reduce it
| Side | Examples | Who controls it | How to reduce it |
|---|---|---|---|
| Oxidiser | Oxygen, nitrous oxide | Anaesthesia | Lowest oxygen that keeps the patient safe, air instead of open oxygen, secure the airway if high oxygen is needed, reduce oxygen before ignition near the airway |
| Ignition | Electrosurgery, laser, fibreoptic light cables, high-speed burrs | Surgery | Bipolar where possible, the lowest effective power, the pencil in its holster when not in use, a warning before use near the airway |
| Fuel | Drapes, gauze, alcohol preparation, hair, the tracheal tube | Nursing and surgery | Let preparation dry, wet gauze and hair near the field, laser-resistant tubes for airway laser |
| Electrical path | Return pad, ECG electrodes, metal touching the skin, pacing wires | The whole team | Good pad contact, no metal touching the patient, insulated pacing wires, a plan for implanted devices |
In practice
- Most fires happen on the head, neck and upper chest under sedation with open oxygen, rather than in the airway itself.
- A pad burn is often found only after surgery, as a burn under the return electrode; check the skin when the pad is removed.
- For a patient with a defibrillator, anti-tachycardia therapy is usually suspended for surgery with monopolar electrosurgery, with external pads ready, and the device is rechecked afterwards.
- Microshock: never touch an exposed pacing wire or the hub of a central line while touching electrical equipment; insulate exposed temporary pacing wires.
- Laser: everyone wears eye protection matched to the laser, and airway laser uses a laser-resistant tube whose cuff is filled with coloured saline so a puncture is seen.
Next step
- Include fire risk in every time-out; when it is high, agree the oxygen plan and who will call stop.
- Know where the gas shut-off, the saline and the extinguisher are in your theatre before the case starts.
- For an implanted cardiac device, obtain the device details and a plan from cardiology or the device clinic before surgery, and have it rechecked afterwards.
- Report every fire, burn or electrical incident through the hospital safety system, and keep the equipment for review.
Common traps
- Mistaken belief: turning the oxygen down just as the surgeon starts is enough. Correction: oxygen lingers under drapes; it must be reduced or stopped with time to clear before ignition near the face or airway.
- Mistaken belief: alcohol preparation is safe as soon as the skin looks dry. Correction: it pools in hair, the umbilicus and under drapes; allow drying time and remove soaked materials.
- Mistaken belief: an airway fire is managed by leaving the tube in and ventilating harder. Correction: the tube is the fuel; remove it at once and stop the gas.
- Mistaken belief: electrical injury always needs a visible shock. Correction: microshock through a central catheter or pacing wire can cause ventricular fibrillation with a current too small to feel.
Test yourself
Think of the answer, then tap the question to see it.
What are the three sides of the fire triangle, and who in theatre controls each?
The oxidiser (oxygen or nitrous oxide), controlled by anaesthesia; the ignition source (electrosurgery, laser, light cables), controlled by surgery; and the fuel (drapes, gauze, skin preparation, tube), controlled mainly by nursing.
Why is a fire most likely during face surgery under sedation?
Open oxygen collects under the drapes next to the face, flammable drapes and preparation are nearby, and electrosurgery is used in the same area: all three sides meet.
What are the first actions in an airway fire?
Stop the gas flow and remove the tracheal tube at once, pour saline into the airway, ventilate with air, then re-intubate and inspect with a bronchoscope.
Why does a poorly attached return pad cause a burn?
The current leaves through a smaller area, so its density and the heat it makes rise enough to burn the skin under the pad.
What is microshock, and how is it prevented?
A small current delivered straight to the heart through a pacing wire or fluid-filled catheter, which can cause ventricular fibrillation; it is prevented by insulating wires and never touching them and equipment at the same time.
Elsewhere in DurgiPedia
Connections
Educational content; it does not replace supervised training, your institution's protocols or the anaesthetist's or intensivist's decision. It contains no doses and no ventilator or infusion settings.
Sources: Surgical Fire Safety (StatPearls (NCBI Bookshelf), 2023); Electrosurgery (StatPearls (NCBI Bookshelf), 2023); Electrical Injuries (StatPearls (NCBI Bookshelf), 2025); Operating Room Fire Safety (Ochsner J, 2011); Electrosurgery and clinical applications of electrosurgical devices in gynecologic procedures (Med J Islam Repub Iran, 2018).
Spotted a mistake?We are always improving for the community. Tell us what is wrong and our team will review it.Report a mistake
Representative model, for educational purposes only.