The anaesthesia machine and breathing circuits, in principle

  • Concept

The anaesthesia machine mixes oxygen, air and anaesthetic vapour in safe proportions and delivers them to the patient through a breathing circuit.

Updated

What it is

The anaesthesia machine takes gases from the hospital pipelines or from cylinders, reduces them to a safe working pressure, mixes them in chosen proportions, adds an exact concentration of volatile anaesthetic and delivers the fresh gas to a breathing circuit connected to the patient. It can be described in three parts. The high-pressure system holds the cylinders, with pin-index connectors so that a cylinder cannot be fitted to the wrong yoke. The intermediate-pressure system takes pipeline gas through gas-specific, non-interchangeable connectors, and includes the oxygen failure alarm, the fail-safe valve that cuts other gases when oxygen pressure falls, and the oxygen flush. The low-pressure system holds the flowmeters, the hypoxic guard that links oxygen and nitrous oxide flows so that a hypoxic mixture cannot be set, and the agent-specific, temperature-compensated vaporisers. The fresh gas reaches the patient through a breathing circuit: most adults are anaesthetised with a circle system, in which exhaled gas passes through a carbon dioxide absorbent and back to the patient through one-way valves, so low fresh gas flows can be used and agent and heat are saved; open circuits without absorbent (Mapleson types and the T-piece used in children) need higher flows to wash out carbon dioxide. The circuit includes a reservoir bag, an adjustable pressure-limiting valve, a ventilator and scavenging of waste gas. A full machine check before each list, and a shorter one before each case, is mandatory.

Key points

  • Gas pipelines and cylinders are colour-coded, but the codes differ between countries: always read the label.
  • Vaporisers are agent-specific and filled with keyed fillers to prevent the wrong agent being added.
  • Desflurane needs a heated, pressurised vaporiser because it boils near room temperature.
  • Scavenging removes exhaled anaesthetic gases so the theatre staff are not exposed to them.
  • Dry absorbent can react with volatile agents and produce carbon monoxide or other harmful compounds; change it as the protocol says.

Components

  • Gas supply: pipelines with wall outlets and backup cylinders on the machine.
  • Pressure regulators, oxygen failure alarm and fail-safe valve.
  • Flowmeters with the hypoxic guard, and the oxygen flush.
  • Vaporisers and the common gas outlet.
  • Breathing circuit: inspiratory and expiratory limbs, one-way valves, absorbent, reservoir bag, adjustable pressure-limiting valve, Y-piece, filter.
  • Ventilator, airway pressure and volume monitors, and scavenging.

How it works

  • Gas flows from high pressure to low pressure: regulators protect the machine and the patient from pipeline and cylinder pressures.
  • Part of the fresh gas passes through the vaporising chamber and picks up saturated vapour; the dial sets how much, so the output concentration is accurate.
  • In the circle, one-way valves keep gas moving in one direction, so exhaled carbon dioxide passes the absorbent before it can be breathed again.
  • The adjustable pressure-limiting valve lets excess gas out to scavenging during spontaneous or manual ventilation; the ventilator takes over this during mechanical ventilation.

Step by step

  1. Check that a self-inflating bag and backup oxygen are present — they are the rescue if the machine fails.
  2. Confirm pipeline connections by tugging and check cylinder contents — a wrong or empty supply is a classic cause of hypoxia.
  3. Check flowmeters, the hypoxic guard and the oxygen analyser calibration — they stop a hypoxic mixture reaching the patient.
  4. Check vaporisers are filled, seated and locked, with no leak — a loose vaporiser leaks agent and gas.
  5. Leak-test the breathing circuit and test the one-way valves and absorbent — leaks and stuck valves cause awareness or rebreathing.
  6. Test the ventilator, alarms, suction and scavenging, and record the check — each is needed during the case.

Machine safety features and what they prevent

FeatureWhat it prevents
Pin-index and gas-specific pipeline connectorsConnecting the wrong gas
Oxygen failure alarm and fail-safe valveDelivering other gases unnoticed when oxygen pressure falls
Hypoxic guard (linked oxygen and nitrous oxide flows)Setting a hypoxic fresh gas mixture
Inspired oxygen analyser with low alarmA hypoxic mixture from any cause reaching the patient
Agent-specific vaporisers with keyed fillers and interlockFilling with the wrong agent or using two vaporisers at once
Adjustable pressure-limiting valve and pressure alarmsExcessive airway pressure and barotrauma

Next step

  • Never start an anaesthetic without a completed machine check and a self-inflating bag within reach.
  • If you suspect machine failure, ventilate the patient with the self-inflating bag and independent oxygen, keep the patient anaesthetised intravenously and call for help.
  • If inspired carbon dioxide rises, check the absorbent and the one-way valves, increase fresh gas flow and change the absorbent.
  • Report any fault to the biomedical engineering team and label the machine out of service until repaired.

Test yourself

Think of the answer, then tap the question to see it.

What does the hypoxic guard on an anaesthesia machine do?

It links oxygen and nitrous oxide flows so that a fresh gas mixture with too little oxygen cannot be set.

How does a circle breathing system remove carbon dioxide?

One-way valves direct exhaled gas through a carbon dioxide absorbent before it returns to the patient, so gas can be rebreathed and low fresh gas flows used.

What should you do if you suspect the anaesthesia machine has failed during a case?

Ventilate with a self-inflating bag on an independent oxygen source, keep the patient anaesthetised intravenously and call for help.

Why is the oxygen flush used with caution during a case?

It delivers pure oxygen at a high flow straight into the circuit, bypassing the vaporisers: it can cause barotrauma and lighten anaesthesia, risking awareness.

Add to my cards (4)

Elsewhere in DurgiPedia

Connections

← All of anaesthesia

Educational content; it does not replace supervised training, your institution's protocols or the anaesthetist's or intensivist's decision. It contains no doses and no ventilator or infusion settings.

Sources: Anesthesia Machine (StatPearls (NCBI Bookshelf), 2022).